본문 바로가기 메뉴바로가기

Papers

A harmonic Bz-based conductivity reconstruction method in with influence of non-transversal current density

https://doi.org/10.1080/17415977.2017.1352587

  • AuthorChang-Ock Lee,Eung Je Woo,전기완
  • JournalInverse Problems in Science and Engineering (1741-5977), 26(6), 811 ~ 833
  • Enrollment typeSCI
  • publication date 20180101
Magnetic resonance electrical impedance tomography (MREIT) is a high-resolution conductivity imaging method utilizing measured magnetic flux density data induced by externally injected currents. Most MREIT image reconstructionmethods including the harmonic Bz algorithm adopt iterative schemes to handle the non-linear relation between conductivity and magnetic flux density. Iterative methods, however, may not guarantee a reliable conductivity reconstruction when the measured magnetic flux density data are contaminated with a significant amount of noise. In this paper, we propose a new image reconstruction method which alleviates the technical difficulties of the iterative harmonic Bz algorithm. It effectively reduces the number of iterations by two at most. To improve the image quality, it incorporates the influence of non-transversal current densities.